Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8000A: CAL ADJ /5100,6666 DATE: 08-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 85 CONFIGURATION: Philips PM 6666 CONFIGURATION: Fluke 5100B ============================================================================= # # Source: # Fluke 8000A instruction manual (PN 347906). # # Compatibility: # MET/CAL 3.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- N J W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP Remove the screw located just below the AC line 1.003 DISP jack. Slide the instrument out of the case. 1.004 DISP Connect the UUT to an ac power source. 1.004 DISP Turn the UUT on. 1.004 DISP Warm-up time: 1 hour. 1.004 DISP Ambient temperature: 22C - 25C. 1.004 DISP Relative humidity: less than 70%. 1.005 HEAD PERIOD ADJUSTMENT 1.006 DISP Connect the PM 6666 Input A to the UUT TP5 and COMMON. 1.007 MESS Adjust R20 for a time period of 100ms (+/-5us). 1.007 MESS Variations of the time period must be <= 15us. 1.008 6666 100mT 2V +D PA S 2W 1.009 DISP Disconnect the PM 6666 from the UUT. 1.010 HEAD ZERO OFFSET ADJUSTMENT 1.011 RNG 200mV 1.012 DISP Connect the 5100B and UUT as follows: 1.012 DISP [32] 5100B OUTPUT HI to UUT V-Ohm 1.012 DISP [32] 5100B OUTPUT LO to UUT COMMON 1.013 5100 0mV S 2W 1.014 DISP Adjust R15 for a UUT display of 00.0+/-00.1V. 1.015 5100 * S 1.016 HEAD TURN-OVER ERROR ADJUSTMENT 1.017 5100 190mV S 2W 1.018 DISP Adjust R23 for a UUT reading of 190.0mV. 1.019 5100 -190mV S 2W 1.020 DISP Adjust R15 for a UUT reading of 190.0+/-0.1mV. 1.021 HEAD RANGE ADJUSTMENTS: DC Volts 1.022 5100 190mV S 2W 1.023 DISP Adjust R23 for a UUT reading of 190.0+/-0.3mV. 1.024 5100 * S 1.025 RNG 2V 1.026 5100 1.9V S 2W 1.027 DISP Adjust R25 for a UUT reading of 1.900+/-0.003V. 1.028 5100 * S 1.029 HEAD RANGE ADJUSTMENTS: Ohms 1.030 RNG 20MZ 1.031 RESF 10MZ S 2W 1.032 DISP Adjust R55 for a UUT reading of 10.00+/-0.11 MOhm. 1.033 RESF * S 1.034 RNG 20kZ 1.035 RESF 10kZ S 2W 1.036 DISP Adjust R12 for a UUT reading of 10.00+/-0.05 kOhm. 1.037 RESF * S 1.038 HEAD RANGE ADJUSTMENTS: High Freq. AC 1.039 RNG 20V AC 1.040 5100 19V 20kH S 2W 1.041 DISP Adjust C3 for a UUT reading of 19.00+/-0.21V. 1.042 END